• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Becton Dickinson & CO (Franklin Lakes) BD Vacutainer® K2 EDTA (K2E) 3.6mg Blood Collection Tubes; Tubes, vacuum sample, with anticoagulant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Becton Dickinson & CO (Franklin Lakes) BD Vacutainer® K2 EDTA (K2E) 3.6mg Blood Collection Tubes; Tubes, vacuum sample, with anticoagulant Back to Search Results
Catalog Number 367841
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd vacutainer® k2 edta (k2e) 3.6mg blood collection tubes, one (1) tube underfilled, requiring recollection.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
There were multiple medical device types reported to be involved.The information for the additional device type is as follows.D.2.A medical device type: jka.D.2.B common device name: tubes, vials, systems, serum separators, blood collection.There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k213670; k231373.Investigation summary: bd had not received any samples or photos for investigation.A total of (b)(4) retained samples were visually inspected, and no issues were identified.Functional tests were performed on (b)(4) retained samples, and all samples were within specification with no issues identified.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: draw volume.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Vacutainer® K2 EDTA (K2E) 3.6mg Blood Collection Tubes
Common Device Name
Tubes, vacuum sample, with anticoagulant
Manufacturer (Section D)
Becton Dickinson & CO (Franklin Lakes)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON, DICKINSON & CO. (Broken Bow)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
+1(201)847-5186
MDR Report Key50915166
Report Number1917413-2026-00910
Device Sequence Number20271951
Product Code GIM
UDI-Device Identifier30382903678410
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
BK230980
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2026
Device Catalogue Number367841
Device Lot Number5101354
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-