Model Number NVRO-27
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Tachycardia (2095)
Date of Event 09/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, a 27mm navitor vision valve was selected for implant using a flexnav delivery system (ds).The patient did not have a history of right bundle branch block (rbbb), left bundle branch block (lbbb) and 1st/2nd atrioventricular block (av) block.The length of the membranous septum was 2.3mm.There was no calcification extending beneath the aortic annular plane in the interventricular septum.The patient had severely impaired left ventricle (lv) function.During the procedure, the valve was able to be implanted successfully at a depth of 5mm on the non-coronary cusp (ncc).Post-implant, the cardiac rhythm changed from sinus rhythm to ventricular tachycardia.The patient received one defibrillation shock to restore the rhythm.Due to the absence of cardiac output chest compressions were continued for approximately one-minute until blood pressure was restored.The physician mentioned that it was not related to the valve performance, but indeed likely related to the patient's past medical history.A temporary pacing wire was placed and remained in the hospital for observation.After a few days, the patient's rhythm recovered subsequently discharged.
 
Additional Manufacturer Narrative
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Navitor Vision
Common Device Name
Aortic Valve, Prosthesis, Percutaneously Delivered
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50915422
Report Number2135147-2026-06562
Device Sequence Number20278222
Product Code NPT
UDI-Device Identifier05415067045799
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P190023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberNVRO-27
Device Catalogue NumberNVRO-27
Device Lot Number11452775
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/21/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-