The account reported finding breather pouch not sealed.Photos were received as a sample to aid in the investigation, and the photos received were of the pack inside the breather pouch; finished good production records and finished good pack build were reviewed.While evaluating the received photos, it was clearly observed the breather pouch was not sealed on the manufacturing side, and the reported issue was confirmed.After further evaluation, the issue was likely caused by a manual error in the sealing process during pack assembly; production and qc managers were notified, feedback will be provided to the build team, and the issue will be monitored for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|