| Model Number |
LifeVest WCD 5100 System |
| Medical Device Problem Codes |
Contamination /Decontamination Problem (2895); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported a battery was damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of battery has been completed.The reported problem (device field normal maintenance and distributor complaint) was confirmed due to the battery latch was sticky and contaminated.The root cause for the contamination was ingress of an unknown liquid.Lifevest patient instructions for use remind and warn patients not to expose the lifevest electronic components to liquids.There was no adverse event that resulted from the damaged battery.
|
| |
|
Search Alerts/Recalls
|