| Model Number |
LA1054096 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information that a trilogy 100 ventilator was returned to a third-party service center for evaluation.There was no patient involvement, and no harm or injury reported.The device was returned in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.On device evaluation, it was reported the device did not power on for a full evaluation and no device information was accessible.There was no evidence of foam degradation in the device.The device was scrapped without further evaluation or repair.
|
| |
|
Additional Manufacturer Narrative
|
|
Device exceeds expected life.Dhr not required.The reported failure of the trilogy ventilator to power on is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|
Search Alerts/Recalls
|