MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; TOWEL,OR,DSP,ST,BLUE,DLX,XR,6/PK,12PK/CS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; TOWEL,OR,DSP,ST,BLUE,DLX,XR,6/PK,12PK/CS Back to Search Results
Catalog Number MDT2168206XR
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "disposable blue towels sluffing onto their equipment.".
 
Additional Manufacturer Narrative
The customer reported observing lint debris from their or towels during use, and the facility reported that blue lint was noted on the angiographic wire during a tavr procedure with no reports of linting into the surgical site and no reports of serious injury or medical intervention.The provided photos were reviewed, the vendor was notified of the issue and sent the photos, and trend analysis was performed by reviewing the past two years of complaint data.Based on the results of the investigation, medline confirmed the reported issue and that the product failed to meet specifications.The vendor noted that the cause of the linting issue was the inherent property of the raw material, recycled cotton, which has short fibers and naturally produces lint, and a review of the complaint information as well as the results of the investigation indicated no need for further action at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TOWEL,OR,DSP,ST,BLUE,DLX,XR,6/PK,12PK/CS
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
urmi parekh
three lakes drive
northfield, IL 60093--275
MDR Report Key50915484
Report Number1417592-2026-52009
Device Sequence Number17485306
Product Code FRL
UDI-Device Identifier10080196778584
UDI-Public10080196778584
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDT2168206XR
Device Lot Number26024100018
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2025
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-