| Model Number |
SC-1232 |
| Medical Device Problem Codes |
Fracture (1260); Migration (4003)
|
| Health Effect - Clinical Codes |
Discomfort (2330); Implant Pain (4561); Bladder Dysfunction (4881)
|
| Date of Event |
06/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patients implantable pulse generator (ipg) has migrated causing pain at the implant site.The patient underwent a revision procedure, upon impedance check, found that one lead was completely fractured with high impedances, physician decided to swap the ipg remove the bad lead and insert a new lead.The patient was doing well postoperatively.The explanted device was not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional suspect medical device components involved in the event:
model: infinion cx lead kit, 50cm,
model number: sc-2317-50,
upn: m365sc2317500,
serial: (b)(6),
batch: 7078682,
udi: (b)(4).
|
| |
|