| Model Number |
URF-V3R |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the uretero-reno videoscope for a separate issue.During inspection of the device, the service repair technician indicates that there was corrosion in the bending section.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.Based on the results of the investigation, a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|