| Model Number |
IntelliVue X3 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported an issue with getting the pulse ox wave and the blue numbers but will show zero.It is unknown if the device was in use on a patient at the time of the event.There was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was sent to philips bench repair.A philips bench repair technician (brt) evaluated the device and found that the x3 has a ¿spo2 equip malf¿ alert, which is due to the x3 having the wrong parameter options installed.Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
| |
|