• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU Medical, Inc CADD Solis HPCA Pump; Pump, infusion, pca

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU Medical, Inc CADD Solis HPCA Pump; Pump, infusion, pca Back to Search Results
Model Number 2110
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device delivered a bolus earlier than programmed.The pump was set to administer a 20 ml bolus of 0.2% ropivacaine approximately 3.5 hours after start, with a continuous rate of 1 ml/hour.The settings were verified by two nurses before initiation.Upon starting, the pump immediately began delivering the bolus, which was stopped by the registered nurse after approximately 2.5 ml had been infused.When restarted, the bolus did not resume and no settings were changed.No patient harm or adverse event was reported.
 
Additional Manufacturer Narrative
No product was returned.The reported issue cannot be confirmed as no product was returned for investigation.If the product is returned, this complaint will be reopened for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD Solis HPCA Pump
Common Device Name
Pump, infusion, pca
Manufacturer (Section D)
ICU Medical, Inc
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
+1(224)706-2300
MDR Report Key50915661
Report Number3012307300-2026-09009
Device Sequence Number14897606
Product Code MEA
UDI-Device Identifier15019517154290
Combination Product (Y/N)N
Initial Reporter StateNA
Initial Reporter CountryAS
PMA/510(K) Number
K170982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2110
Device Catalogue Number21-2111-0402-51
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/04/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-