| Model Number |
2110 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device delivered a bolus earlier than programmed.The pump was set to administer a 20 ml bolus of 0.2% ropivacaine approximately 3.5 hours after start, with a continuous rate of 1 ml/hour.The settings were verified by two nurses before initiation.Upon starting, the pump immediately began delivering the bolus, which was stopped by the registered nurse after approximately 2.5 ml had been infused.When restarted, the bolus did not resume and no settings were changed.No patient harm or adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
No product was returned.The reported issue cannot be confirmed as no product was returned for investigation.If the product is returned, this complaint will be reopened for further investigation.
|
| |
|