U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Tandem Diabetes Care Tandem Mobi Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Tandem Diabetes Care Tandem Mobi Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number 1004000
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that pump displayed undertravel stall malfunction alarm, and blood glucose meter showed high value above meter range.Customer¿s blood glucose level exceeded 500 mg/dl, with exact value unknown.Customer gave correction insulin injection using multiple daily injections, switched to backup multiple daily injection therapy, and did not seek help from a third party.Technical support arranged replacement pump with updated software, confirmed shipping details, provided instructions for storage mode and pump return, and advised customer to enter pump settings and reconnect continuous glucose monitor on replacement pump.
 
Additional Manufacturer Narrative
In use at the time of the event: cgm model: unknown.Mobile os: android / android_galaxy s20 fe 5g / android 16.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Tandem Mobi Insulin Pump with Control-IQ+ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50915759
Report Number3013756811-2026-175835
Device Sequence Number14788527
Product Code QFG
UDI-Device Identifier00389152370019
UDI-Public(01)00389152370019(22)1013701(11)240529
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K240309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1004000
Device Catalogue Number1013700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/29/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight195 LB
-
-