The account reported finding adhesive issue with dressing, and the complaint transcript states, "dressing not stay adhered to skin." a sample was received for investigation; medline received 5 ea unopened kit,iv start,chloraprep kits, tested all 5 of the dressings returned, and all of them peeled correctly and the paper frame stayed attached to the dressing, with no defects or damage observed.Finished good production records and finished good pack build were reviewed, and no issues were identified that might have had an impact on or likely contributed to the reported issue.Medline was unable to confirm the reported issue with the samples received, was unable to determine the root cause with the dressings that were received, and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|