• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBQ-CEDIV ADMIN SET, 15DRP, SECOND, 40

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBQ-CEDIV ADMIN SET, 15DRP, SECOND, 40 Back to Search Results
Catalog Number DYNDTB1540
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "one was spiked into an acetaminophen bottle and when the tab was opened to allow in air in, it started leaking ¿ a lot.It was connected to the primary tubing that was hooked up to an iv pump.".
 
Additional Manufacturer Narrative
The account reported finding leaking; the customer reported the secondary tubing was spiked into an acetaminophen bottle and when the tab was opened to allow in air in, it started leaking, and it was connected to the primary tubing that was hooked up to an iv pump.No sample or photo received for our investigation; finished good production records were reviewed, the manufacturer was notified of this incident and requested a completed review of the production records, no investigation was received from the manufacturer after two follow ups, and component trending was reviewed with 0 additional reported complaints for this issue for component dyndtb1540 in the last 24 months.Based on the complaint details and our investigation, the complaint is not confirmed; at this time, we are unable to confirm the reported defect, as no physical sample or detailed photographic evidence was provided to support the investigation.Sample required to determine cause; this complaint has been provided to the vendor for further evaluation, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBQ-CEDIV ADMIN SET, 15DRP, SECOND, 40
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
daniel soto
three lakes drive
northfield, IL 60093--275
MDR Report Key50915974
Report Number1417592-2026-52071
Device Sequence Number5330484
Product Code FPA
UDI-Device Identifier10197344098637
UDI-Public10197344098637
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K210381
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNDTB1540
Device Lot NumberBQ241052101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-