• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; LIFT,PATIENT,STAND ASSIST,ELECTRC,40

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; LIFT,PATIENT,STAND ASSIST,ELECTRC,40 Back to Search Results
Catalog Number MDS400SAB
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the wheel has literally split in half and broke apart.".
 
Additional Manufacturer Narrative
It was reported that customer stated that the wheel is fractured in half; the customer later stated, "the piece is not what came with the lift, that looked more like a key ring, and those have broken," and stated there have been no resident injuries or falls.Based on the photo evidence provided by the customer, the product appears to be in used condition due to the wear on the caster; visual review found the plastic of one locking caster has fractured and appears to be splitting in half off of the caster, and additional photos showed the customer has replaced the circular rings attached to the hooks on the 4-point boom with rectangular shaped rings.The investigation results were confirmed.The root cause of this issue with the caster is unknown, and the ring issue would be an issue of customer use/misuse; no trends were identified, no product changes were identified that could have contributed to the alleged product failure, and the division will monitor complaints through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
LIFT,PATIENT,STAND ASSIST,ELECTRC,40
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rahul kota
three lakes drive
northfield, IL 60093--275
MDR Report Key50915991
Report Number1417592-2026-52074
Device Sequence Number17478942
Product Code FSA
UDI-Device Identifier10193489019254
UDI-Public10193489019254
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS400SAB
Device Lot Number68722042597
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2025
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-