It was reported that "the assembly process has been an inconvenience for the team, but more concerning is that they fall apart during examination.Additionally, the staff has noted significant muffling during auscultation, which makes it difficult to properly assess patients' cardiac status.".
|
|
The customer reported receiving medline disposable yellow stethoscopes that were falling apart during use, and reported that "the assembly process has been an inconvenience for the team" and staff noted "significant muffling during auscultation." medline received 20 stethoscopes in new condition and randomly selected 5 stethoscopes for visual and functional testing; no manufacturing defects preventing the samples from working as intended were seen, the stethoscopes were assembled according to the manual, and a functional test was performed.During the testing process, there were no issues with the samples falling apart; they remained fully assembled, even when excessive force was applied to the tubing, the samples operated as intended, and the complaint was not confirmed.The cause was not confirmed with the received sample, and based on the review of the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|