• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; TRAY,TRACH,SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; TRAY,TRACH,SALINE Back to Search Results
Catalog Number DYND40589
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "in the tracheostomy clean and care kit; dynd40589, the bristles in the tracheostomy brush are falling off.".
 
Additional Manufacturer Narrative
The account reported finding "bristles falling off" and the customer reported that the bristles in the tracheostomy brush are falling off.Medline quality received one case of (b)(6); the sample arrived in its original packaging, and a portion of the received samples were sent to the manufacturer for investigation.Based on the manufacturer's investigation, the bristles were not falling off at the connection to the handle; pieces of the bristle broke off during a phase of the manufacturing process where the bristles are attached to the handle.The root cause was determined to be the manufacturing process, and the manufacturer has installed corrective actions to better detect these occurrences during the manufacturing/packaging process.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TRAY,TRACH,SALINE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
vamshi thupakula
three lakes drive
northfield, IL 60093--275
MDR Report Key50916073
Report Number1417592-2026-52098
Device Sequence Number17642010
Product Code LRO
UDI-Device Identifier40080196032823
UDI-Public40080196032823
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND40589
Device Lot Number2024031390
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/02/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2024
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-