| Model Number |
LSP203A |
| Medical Device Problem Codes |
Stretched (1601); Use of Device Problem (1670); Capturing Problem (2891)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient presented for an initial implant procedure of an atrial leadless pacemaker (alp) on (b)(6) 2026.Upon fixation, the current of injury and capture thresholds were unacceptable and the physician elected to reposition the alp at which time it was noted the helix stretched.The alp was not used, and a new alp was successfully implanted.The patient was stable throughout the procedure.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated the alp lost capture at 0.1 milliseconds (ms) pulse width and had intermittent capture at 1.5 ms pulse width.While trying to retrieve the alp to reposition it, it was noted the alp would not dock.Advance maneuvers were used and the entire system was retrieved without issue.
|
| |
|