| Catalog Number |
XA23LS60ZSC081N |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tracheostomy tube experienced unintended inflation of the inner cuff when air was introduced into the outer cuff, which prevented the patient from speaking.The issue was identified after the device had been placed in the patient's airway.There was patient involvement associated with this event; however, no patient harm or adverse effects were reported, as the patient remained under continuous monitoring.
|
| |
|
Additional Manufacturer Narrative
|
|
This file was originally incorrectly filed under registration number (b)(4).The date of that submission was 27-may-2026.One sample was returned for evaluation.A review of the lot history identified no nonconformities that could have contributed to the reported complaint.Visual inspection and functional testing were performed, and no anomalies were observed.The reported issue could not be replicated, as inflation or deflation of one cuff did not influence the other cuff during testing.Therefore, the complaint was not confirmed.
|
| |
|