The mayfield composite series base unit, standard (a3101) was returned for evaluation:
device history record (dhr) review - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - inspection of the mayfield base unit a3101 ((b)(6)) determined that the screw on the standard adapter a3018 exhibited damaged threads and could not be properly threaded into the knob of the connecting element.The damaged screw on the standard adapter a3018 was replaced, restoring proper engagement with the connecting element knob.Following completion of the repair, the device underwent functional testing and successfully met all applicable performance requirements.Root cause - the complaint is confirmed via inspection of the unit.The screw on the a3018 adapter had damaged threads and could not be screwed into the knob of the transitional (connecting element).The probable root cause is damage from mishandling of the unit.Additionally, improper or suboptimal placement of the mayfield system can contribute to movement of the unit or patient.Based on the evaluation of the event, no further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|