This vmsr summary report covers malfunction events that occurred in q3 of 2025.The report summarizes 8 malfunction events involving sphygmomanometers (mds9380).Customer reported that "cuffs are defective, the inflation bulb is not working it gets stuck together while taking blood presure." associated serial numbers: (b)(6).Associated di: (b)(4).Lot number 65122030002, 65124090002, and 65125020002.Medline received 8 samples, tested each sample for inflation to 300mmhg, and switched bulbs to other cuffs to test that the bulbs worked once removed and switched.All bulbs were able to inflate multiple cuffs and medline did not experience any bulb getting stuck while inflating; however, one cuff did experience leakage.The reported bulb issue was not confirmed, one cuff leakage was confirmed, and the cause was identified as comp./part defective, broke; there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|