(b)(4)
date sent: 9/28/2026
d4 batch#: a97v33.Additional information was requested and the following was obtained:
was surgery delayed due to the reported event? yes,
if yes, number of minutes: 20,
was procedure successfully completed? yes
were fragments generated? no
if yes, were they removed easily without additional intervention? no
patient status/ outcome / consequences no
was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: no
investigation summary
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, with evidence of contact with metal in or out of the operative field.During functional testing on energy systems, an alert screen was displayed.A probable cause for the device to stop activating and the energy systems to display an alert screen is blade damage.The device was analyzed, and it was determined that it was possibly used in more than one procedure, based on generator data.The device is intended and labeled for single patient use.If the device is connected to multiple generators, an alert screen will be displayed indicating ¿adaptive tissue technology features are not available in this device¿.Due to the multiple generator usage, which is evidence of possible reuse, we are unable to determine how this condition impacted the performance of the device and, therefore, cannot conclude the root cause.No communication issues were observed at any time during the testing.The device was disassembled to verify the condition of the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.This device is packaged and sterilized for single use only.Multiple patients uses may compromise the device's integrity or create a risk of contamination that, in turn, may result in patient injury or illness.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch a97v33, and no non-conformances were identified.
|