Catalog Number HAR36
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure the device shows multiple errors "replace instruments" and "identifying device".Completed the procedure with the help of another device.No patient consequences reported.
 
Additional Manufacturer Narrative
(b)(4) date sent: 9/28/2026 d4 batch#: a97v33.Additional information was requested and the following was obtained: was surgery delayed due to the reported event? yes, if yes, number of minutes: 20, was procedure successfully completed? yes were fragments generated? no if yes, were they removed easily without additional intervention? no patient status/ outcome / consequences no was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: no investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, with evidence of contact with metal in or out of the operative field.During functional testing on energy systems, an alert screen was displayed.A probable cause for the device to stop activating and the energy systems to display an alert screen is blade damage.The device was analyzed, and it was determined that it was possibly used in more than one procedure, based on generator data.The device is intended and labeled for single patient use.If the device is connected to multiple generators, an alert screen will be displayed indicating ¿adaptive tissue technology features are not available in this device¿.Due to the multiple generator usage, which is evidence of possible reuse, we are unable to determine how this condition impacted the performance of the device and, therefore, cannot conclude the root cause.No communication issues were observed at any time during the testing.The device was disassembled to verify the condition of the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.This device is packaged and sterilized for single use only.Multiple patients uses may compromise the device's integrity or create a risk of contamination that, in turn, may result in patient injury or illness.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch a97v33, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Harmonic ACE
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50916665
Report Number3005075853-2026-08027
Device Sequence Number5803754
Product Code LFL
UDI-Device Identifier10705036014447
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K120729
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberHAR36
Device Lot NumberA97V33
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-