It was reported that during a right upper lobe wedge resection, after an initial firing and loading a reload, the stapler displayed zero reloads.A reinforced reload was then used and fired; however, the knife blade failed to retract, and the reload became lockedon tissue and could not be removed.Attempts to open the reload branches with a screwdriver were unsuccessful.Another reload was used to re-staple beneath the stuck staple line, and additional right upper lobe tissue was resected.There was no patient injury.Pli50: product was received without accompanying information.
|
|
D10 concomitant products: egia45amt, egia45amt egia 45 artic med thick sulu, (lot # unknown) sigphandle, sig power sigphandle handle, (serial # c24aaa0649) sigadaptshort, sig power sigadaptshort lin short adapt, (serial # (b)(6)) sigpshell, sig power sigpshell control shell, (lot # unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|