• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Beckman Coulter, Inc Access High Sensitivity Troponin I reagent; Immunoassay Method, Troponin Subunit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Beckman Coulter, Inc Access High Sensitivity Troponin I reagent; Immunoassay Method, Troponin Subunit Back to Search Results
Catalog Number B52699
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6)2024, the customer reported erroneously elevated hstni patient results.There was no report of an injury or illness to the patient attributable to the output from the device in this event.The customer reported the following hstni results with two patient samples.Units of measure were not provided: initial result 729 repeat on same and alternate instrument 8-11 initial result 83 repeat 8.29 repeated on alternate instrument 6 three levels of quality controls (qc) were passing within the laboratory¿s established ranges the customer reported both samples were lithium heparin and one sample was not a full draw and customer unsure about sample mixing and sample was centrifuged on the line.No hardware errors were reported in conjunction with this event.Fse was dispatched to the customer's site.
 
Additional Manufacturer Narrative
A1: the full patient identifier is (b)(4).A2, a3, a4, a5 and a6: the customer did not supply patient demographics such as age, date of birth, sex, weight, ethnicity or race.H3 and h6: no hardware errors or other sample/assay issues were reported in conjunction with this event.System performance indicators such as system check, calibration and quality control were passing within specifications at the time of the event.The customer reported both samples were lithium heparin and one sample was not a full draw and customer unsure about sample mixing and sample was centrifuged on the line.According to access hstni instruction for use (ifu), part number c09448, it states: ¿¿the role of pre-analytical factors in laboratory testing has been described in a variety of published literature.Following blood collection tube manufacturers¿ specimen collection and handling recommendations are essential to reduce pre-analytical errors¿.Additionally, as per clsi "[clinical and laboratory standards institute]" guideline-gp44-a4: procedures for the handling and processing of blood specimens for common laboratory tests.4th edition, ¿if less blood than required is drawn, the excess amount of additive has the potential to adversely affect the accuracy of test results¿ a fse was dispatched to the customer's site.Field service engineer (fse) verified instrument hardware by running a system check, high sensitivity system check, carryover and low-level qc precision and reported all passed.No further issue was reported by the customer.In conclusion, the cause of the event is use-error as customer did not follow the sample handling instructions for use.Retrospective mdr submission under capa-001570.The complaint was reassessed using beckman coulter's revised mdr malfunction reportability methodology and was determined to be reportable.This report is being submitted as part of capa-001570 remediation activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Access High Sensitivity Troponin I reagent
Common Device Name
Immunoassay Method, Troponin Subunit
Manufacturer (Section D)
Beckman Coulter, Inc
1000 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
Beckman Coulter, Inc
1000 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
shama agtey
1000 lake hazeltine drive
chaska, MN 55318
+1(952)491-6938
MDR Report Key50916694
Report Number2122870-2026-00339
Device Sequence Number5804246
Product Code MMI
UDI-Device Identifier15099590693183
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K172787
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Catalogue NumberB52699
Device Lot Number440081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2024
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-