The customer reported that the aneroid mobile adult cuff "does not work properly" and "the readings a=e not accurate." medline received 1ea mobile aneroid blood pressure monitor in like-new condition; visual inspection showed the dial was off from the start position, the unit was calibrated by inserting a calibration tool into the stem where the cuff tubbing is inserted, the gauge was adjusted to the initial position, and the cuff was inflated up to 300 mmhg five times with no issues experienced.The complaint was not confirmed; the cuff inflated/deflated normal, the gauge displayed no issues, overall the sample worked as intended, and material trending via sap showed no similar complaints during the last 12 months.The cause code was not confirmed with received sample, and the division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|