Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited a gradual rise to out of range shock impedance measurements.Technical services (ts) recommended bringing the patient into the clinic to clear the alert and increase the alert value to 150 ohms.Ts also recommended continued monitoring for the patient.No adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|