Catalog Number SNAPSECURE3W
Medical Device Problem Codes Break (1069); Material Deformation (2976); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "top plastic part is snapping off, some not clamping at all, teeth notch failing.".
 
Additional Manufacturer Narrative
The reporter described snapsecure latch and hinge failures, including that the top plastic part is snapping off, some not clamping at all, teeth notch failing, and the teeth plastic piece folds inward and won¿t line up to click in and easily breaks off totally.The complaint stated no patient was harmed and no medical intervention was necessary; a new cathgrip was used in its place and the securement was replaced.Samples were sent and a sample assessment and quality investigation was completed.The complaint file lists the cause as unconfirmed defect and sample required to determine cause; these issues have been addressed with the account, medline urology qa will continue to monitor this issue, and no further action is needed at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
SNAPSECURE 3-WAY FOLEY DEVICE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093
Manufacturer Contact
william abrams
three lakes drive
northfield, IL 60093
MDR Report Key50916896
Report Number1417592-2026-52219
Device Sequence Number14897490
Product Code EYJ
UDI-Device Identifier20193489102717
UDI-Public20193489102717
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSNAPSECURE3W
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2024
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-