Model Number BI70002000
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding an imaging system being used during a cervical spine procedure.It was reported that during a cervical spine case they felt they were 5mm off (inferior) after the spin.They swapped the frame clamp and they were still off by 5mm (inferior) even after re-spinning with a new clamp (alleged inaccuracy). as they worked towards the frame they found that accuracy improved, but was never perfect.Additional information received "resolution confirmed on call". this system had since been used with an axis and was seemed to be accurate".
 
Additional Manufacturer Narrative
(h3,h6) no parts have been received by the manufacturer for evaluation.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi-500-01145, version #:4.2.0, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-Arm O2 Imaging System
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (Boxborough)
200 beaver brook rd
boxborough,ma 45640
MX  45640
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (Boxborough)
200 beaver brook rd
boxborough,ma 45640
MX   45640
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50916958
Report Number3004785967-2026-00520
Device Sequence Number5394641
Product Code OWB
UDI-Device Identifier00763000132378
UDI-Public(01)00763000132378(11)190205
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K200074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Patient SexMale
Patient Weight195 LB
-
-