| Model Number |
WA50042A |
| Medical Device Problem Codes |
Leak/Splash (1354); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the gynecologic laparoscope¿to the service center for evaluation related to a separate issue.During testing, the service repair technician identified dirt on the grip and water leakage in the device.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction dirty grip was water leakage into the device.Additionally, the malfunction was traced to a component failure of the main body.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|