| Model Number |
PCF-H190L |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported an image that was all staticky, like snow during inspection for use involving an olympus colonovideoscope.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection, and the reported failure was confirmed.Based on the results of the investigation, the cause was a damaged charge coupled device unit.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|