• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Inc. STRATAFIX; Suture, surgical, absorbable, polydioxanone

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Inc. STRATAFIX; Suture, surgical, absorbable, polydioxanone Back to Search Results
Catalog Number SXPP1B112
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a patient underwent an unknown surgery on (b)(6) 2026 and barbed suture was used.The suture was broken after passing it through just one stitch when used for wound closure.It was not a control release needle.Another product was used to complete the case.There were no adverse consequences to the patient.Further details are not provided.Additional information has been requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.What is the current status of the patient? =>unk.-when was the suture broke (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify=>during use on the patient - could you kindly perform and document the follow-up attempt for the product return? =>the device has been received at sukagawa and will be shipped.Please check rmao.- please provide the source or name of person providing answers to follow-up questions: (b)(4).H3 evaluation: the product was returned to eth for evaluation.Visual inspection revealed that one one dispensed needle suture combination and one empty labeled winding former were received for analysis.Product code sxpp1b112.During visual inspection of the sample, the swage and attachment area were noted to be as expected.The suture was examined, and it was observed that the weld of the first loop was detached.Besides that, body fluids were found on the strand.Per the condition of the returned sample, it is possible that this failure was due to excessive force being applied.As part of the eth quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as on what caused the reported complaint.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
Suture, surgical, absorbable, polydioxanone
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.-San Lorenzo PR
road 183, km 8.3
san lorenzo 00754
PR   00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50917326
Report Number2210968-2026-09769
Device Sequence Number5394106
Product Code NEW
UDI-Device Identifier10705031235984
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K182873
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1B112
Device Lot Number107T8U
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/22/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-