Product complaint # (b)(4).Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.What is the current status of the patient? =>unk.-when was the suture broke (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify=>during use on the patient
- could you kindly perform and document the follow-up attempt for the product return? =>the device has been received at sukagawa and will be shipped.Please check rmao.- please provide the source or name of person providing answers to follow-up questions: (b)(4).H3 evaluation:
the product was returned to eth for evaluation.Visual inspection revealed that one one dispensed needle suture combination and one empty labeled winding former were received for analysis.Product code sxpp1b112.During visual inspection of the sample, the swage and attachment area were noted to be as expected.The suture was examined, and it was observed that the weld of the first loop was detached.Besides that, body fluids were found on the strand.Per the condition of the returned sample, it is possible that this failure was due to excessive force being applied.As part of the eth quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as on what caused the reported complaint.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|