The reported issue was the transducer cracked during the setup of procedure.A review of the device history records was performed for the indicated packaging and component lots, and the sample was reviewed.Based on the sample supplied from the customer, the customer's complaint of "bond break between transducer and one way stopcock" is deemed confirmed; possible root causes may be due to the end user trying to tighten the bonded connection or handling damage during transit.There does not appear to be any manufacturing defects related to the returned bonded assembly, and no corrective action is needed at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|