Product complaint # (b)(4).Date sent to the fda: 9/28/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained:
could you please confirm the lot number? (the number provided is incomplete) the lot number for the suture is lot# 10eabx.I pulled the box from the shelf and i am returning the entire box since the broken needle was not kept by the hospital ¿ it was put into a sharps container by mistake instead of bagged in a biohazard bag requested by the surgeon.The hospital is requesting a replacement box.Did the needle fall into the patient? according to the surgeon the broken piece was found in the patients¿ fascia.It did not fall but was caught in the patient¿s tissue
if yes, was the needle piece(s) retrieved during the same procedure? the needle piece was retrieved ¿ i have a photo of the 2 pieces
were x-rays taken to locate the needle piece(s)? yes, x-rays were taken since it was not apparent where the broken piece was located on the patient
what measures were implemented to retrieve the broken piece? after the piece was located on the patient the broken piece of needle was carefully extracted from the patient¿s fascia
was there any additional tissue damage resulting from the search for the needle piece? the patient incision was extended to assist in retrieving the broken needle
does a fragment of the needle remain in the patient¿s tissue? no, the fragment did not remain in the patient.If yes, was more than half the needle retained in the patient? n/a
is there any plan in place to remove the needle piece in the future? n/a
if yes, could you please provide the scheduled date for the removal? n/a
in which tissue structure was the broken needle located? located in the patients¿ fascia
what is the surgeon's opinion of the potential consequences for the patient? dr.(b)(6)(please refer to the attached email) said there were no potential consequences for the patient.She confirmed the patient was doing well.Please provide the source or title of external person providing answers to follow-up:
(b)(6).¿ please provide picture mentioned if available
¿ please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep)
attached is the photo of the needle.I had a follow-up conversation with dr (b)(6) who was the surgeon using the needle that broken during the case
to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A photo has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|