| Catalog Number |
08836973190 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys hiv duo results for 1 patient sample on a cobas e 801 analytical unit.The initial result was 73.3 coi (positive).A western blot test and an hiv viral load test were performed on the patient sample and the results were negative.
|
| |
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The calibration and qc recovery data provided was within specification.The patient sample was received for investigation.The investigation produced a reactive result with the provided patient sample.But a non-reactive result with the elecsys hiv duo ii reagent.False reactive results may occur in elecsys hiv duo as the assay has a sensitivity of <100%.The elecsys hiv duo assay performs within specification.No product problem was identified.
|
| |
|