| Model Number |
CF-Q150L |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the banding in the image was displayed involving an olympus colonovideoscope.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.\
\
updated fields include d9, h2, h3, h6, and h11.\
\
the final investigation has been completed.The failure reported by the customer was confirmed.\
\
the most likely cause of the reported issue was traced to slipping in the charge-coupled device (ccd) unit.\
\
should additional relevant information become available, a supplemental report will be submitted.\
\
olympus will continue to monitor the field performance of this device and evaluate any additional information received through post market surveillance activities.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.\
\
the investigation is currently ongoing.A supplemental report will be submitted upon completion of the investigation or should new and significant information become available.
|
| |
|