It was reported that this was an arteriotomy closure using a starclose device after a procedure.Reportedly, after deploying the clip, the flex guide or delivery tube snapped "[audible noise]", forcing the physician to pull the clip out of artery "[clip stuck on distal end of sheath or device when device was withdrawn]".Manual compression was used to achieve hemostasis.There were no reported adverse patient effects.No additional information was provided.
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot of specific issue.Based on the information provided, a definitive cause for the reported issue could not be determined.Factors that contribute to firing problem and audible sound of the device may include, but are not limited to, inadequate nick and spread, user presses the firing button (#4) prior to full advancement of the thumb advancer, user pulls back on device during clip deployment.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.B3 - estimated date.
|