Analysis was performed remotely by service personnel and included remote access to the primary server, access to the involved computer authorized by biomed, review of sound settings under system configuration > tools > control panel, and efforts to reproduce the issue.Remote service was able replicate the reported issue.A review of the configuration settings found that the speaker was not selected as the default audio device.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was a configuration issue.The issue was resolved by configuring the audio speaker.The biomed provided confirmation that the issue was resolved.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|