Olympus Winter & Ibe GmbH Video telescope "ENDOEYE HD II", 10 mm, 30°, autoclavable; Laparoscope, Gynecologic
|
Back to Search Results |
|
| Model Number |
WA50042A |
| Medical Device Problem Codes |
Break (1069); Degraded (1153); Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported image discoloration during an unspecified procedure involving an olympus laparoscope, gynecologic, which was completed with the same device.Additionally, during the initial evaluation of the device, a broken charge coupled device unit, distal end of the insertion section with contour sharpness, and a green image were observed.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation regarding the allegation and the green image, the cause was a broken charge coupled device unit.For the distal end of the insertion section with contour sharpness, the cause was traced to a component failure without any design or manufacturing issues.The date of the event is unknown.Additional information will be provided if available.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|