| Model Number |
774F75 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during use, a swan ganz catheter had deflation difficulties.Balloon was tested prior to insertion without issue.Balloon was successfully floated however there was no passive deflation when instructed to deflate.Syringe was exchanged for a different syringe, but there was still no deflation and resistance was felt.Swan was kept in the patient who was then transferred to intensive care unit.There was no patient injuries.Further information is unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional product code: dqo, dqe, kra.This device is not available for return.Without the return of the unit, it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.A device history record review was completed and no related non-conformances were found.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.Full udi: (b)(4).
|
| |
|