|
It was reported that, during video-assisted thoracic surgery for the lung resection, the device could not activate and the knife did not move when the tissue was grasped and fired.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.R.
|
|
(b)(4).Date sent: 9/28/2026.D4: batch #a9m559.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ech60s device was returned with no apparent damage and with no reload present.The device was returned with a non-working firing mechanism.No further functional test could be performed due to the condition of the device.In order to verify the condition of the internal components, the device was disassembled.Upon disassembling, the pcb was noted to be non-functional.No conclusion could be reached as to what may have caused the pcb to be non-functional.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number a9m559, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|