MEDTRONIC PUERTO RICO OPERATIONS CO. SynchroMed III; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8667-40
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a patient regarding an external device to an implantable pump infusing unknown baclofen for intractable spasticity.It was reported that the past couple nights ((b)(6) 2026) the patient had been trying to bolus and the controller kept saying 0% and did not do anything.The patient stated they turned the controller off and back on again and it acted like it gave them a bolus but they could not tell if it gave the bolus or not.The agent walked the patient through clearing the data.The patient was able to connect and then see the bolus screen.
 
Additional Manufacturer Narrative
Continuation of d10:product type product id a820 lot# serial# unknown implanted: explanted: product type software section d information references the main component of the system.Other relevant device(s) are: product id: a820, serial/lot #: unknown, ubd: , udi#: medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SynchroMed III
Common Device Name
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50921028
Report Number3004209178-2026-16446
Device Sequence Number5394213
Product Code LKK
UDI-Device Identifier00763000999421
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P860004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Notification,Recall
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8667-40
Device Catalogue Number8667-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2026
Initial Report FDA Received Date09/28/2026
Date Device Manufactured05/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2182207-01-24-2025-001-C
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-