(b)(4).Date sent: 9/28/2026.D4: batch #807d24.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ec45a device was returned with no apparent damage and with a cecr45w partially fired 1/10 reload loaded on the device.The device was returned with the jaws and closure trigger in the closed position.Also, the knife was noted partially advanced, the red manual knife reverse button was activated and the knife returned to the home position as expected.The returned device and reload were tested for functionality in the articulated position by resetting and reloading it into the device.The device achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment slot resulting in the knife pushing the one piece sled forward and locking the reload.As part of ethicon ¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 807d24, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device lot/batch number, a9m743, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|