• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/16/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiv ing fentanyl and unknown morphine drug via an implantable pump for spinal pain indications.It was reported that they made a medication change and they were about half way through the bridge bolus.Today they intended to clear out all the medication by aspirating through the cap but they were only able to get a small amount of medication and were not able to clear the catheter.Caller then stated that patient became narcotized just after the procedure.Caller said that they reversed it with a little narcan but they would like to pause the bridge boluses.Agent reviewed with the caller that they could program the pump to minimum rate mode and resume therapy when they want then complete a modified bridge bolus for the remained of the old drug.Caller stated that the original bolus was 42h and 44min and he is getting ready to stop it now at 22h and 6 min remaining.The healthcare provider (hcp) will call back for further assistance when they decide to turn the pump back.The physician called back stating that they will be restarting the pump.
 
Additional Manufacturer Narrative
Section d references the main component of the system.Other medical products in use during the event include: brand name ascenda; product id 8780 (serial: hg6vb4f16); product type: 0184-catheter; implant date 2023-06-02; explant date medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Common Device Name
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50921456
Report Number3004209178-2026-16447
Device Sequence Number5369945
Product Code LKK
UDI-Device Identifier00763000634100
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P860004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2024
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2026
Initial Report FDA Received Date09/28/2026
Date Device Manufactured05/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexFemale
-
-