| Model Number |
8637-40 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiv ing fentanyl and unknown morphine drug via an implantable pump for spinal pain indications.It was reported that they made a medication change and they were about half way through the bridge bolus.Today they intended to clear out all the medication by aspirating through the cap but they were only able to get a small amount of medication and were not able to clear the catheter.Caller then stated that patient became narcotized just after the procedure.Caller said that they reversed it with a little narcan but they would like to pause the bridge boluses.Agent reviewed with the caller that they could program the pump to minimum rate mode and resume therapy when they want then complete a modified bridge bolus for the remained of the old drug.Caller stated that the original bolus was 42h and 44min and he is getting ready to stop it now at 22h and 6 min remaining.The healthcare provider (hcp) will call back for further assistance when they decide to turn the pump back.The physician called back stating that they will be restarting the pump.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d references the main component of the system.Other medical products in use during the event include: brand name ascenda; product id 8780 (serial: hg6vb4f16); product type: 0184-catheter; implant date 2023-06-02; explant date medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|