Product complaint # (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained:
- please confirm if there is an issue with the applier? no issues.- what suture type and size was used? 3-0 stratafix or vloc.- when the event occurred, was the suture placed near the hinge of the clip? yes.- were you able to lock the clip closed on the suture? if yes, after it closed, was the clip holding securely fixed on the suture? eventually yes and yes.- was the applier checked for damaged (jaws straight and aligned)? no damage to clip applier.- if the clip did not close/hold on the suture, was the clip used in an application where the suture was under tension? the clip was not holding when loading until the tech got it to lock eventually.Yes to tension.Device history review: a manufacturing record evaluation was performed for the finished device 104p1h/xc20022 and no non conformances / manufacturing irregularities related to the malfunction were identified.Mfg.Date: 11/12/2024 expiry date: 10/31/2027.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|