| Catalog Number |
367364 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Report 7 of 8.It was reported while using bd vacutainer® ultratouch¿ push button blood collection set, blood leaked from a crack in the hub of one (1) device.A new device was selected and the patient sample was recollected.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information:
d.9: device available for evaluation: yes
h.3: device eval by manufacturer? yes
d.9: returned to manufacturer on: 28-sep-2026.Investigation summary:
bd received 150 samples for investigation.Out of these, 20 samples were randomly selected and visually inspected; none of the samples indicated a split female luer adapter.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.While bd was unable to confirm this complaint for the indicated failure mode damaged, bd has initiated further root cause investigation relating to the issue of damaged/broken female luer through corrective and preventive actions.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|