It was reported that the ilet displayed alert 57 and the user restarted the ilet; the user also had backup therapy available and was using a dexcom g7 sensor, which aligns with a computer software problem related to the device¿s computer software component.Symptoms included no clinical signs, symptoms, or conditions.Outcomes included no health consequences or impact.Investigation included communication with the user, analysis of information provided by the user or third party, and an event history log review.Investigation of this case revealed a software runtime error consistent with a computer software problem in the computer software component.It was concluded, based on previously established findings for similar reports, that the cause was traced to software coding.If the device is returned, a physical evaluation will be performed, and a supplemental will be submitted.
|