• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HeartWare, Inc. HeartWare Ventricular Assist System - Pump; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HeartWare, Inc. HeartWare Ventricular Assist System - Pump; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1103
Medical Device Problem Code Pumping Problem (3016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the ventricular assist device (vad) patient exhibited a high-watt alarm while operating at the higher end of the expected power range for the programmed speed.The logfiles demonstrated a circadian rhythm that appeared to have nothing adverse showing.The patient¿s urine remains normal in color, with lactate dehydrogenase (ldh) 245 and international normalized ratio (inr) 2.4.They are planning to increase his high watt alarm to 9.0w, they did a speed increase to 9/8 and think it's average power runs high normally because he is a very active young patient.The vad remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HeartWare Ventricular Assist System - Pump
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HeartWare, Inc.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HeartWare, Inc.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50922876
Report Number3007042319-2026-01466
Device Sequence Number5369987
Product Code DSQ
UDI-Device Identifier00888707003261
UDI-Public(01)00888707003261(17)210531
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2021
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2026
Supplement Date Received by Manufacturer10/01/2026
Initial Report FDA Received Date09/28/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age19 YR
Patient SexMale
-
-