The customer reported that they opened a new box of solutions and it was not reading within range.Medline received one evencare g2 low control solutions in used condition and ran a control test using both high and low solutions; the results were 57 mg/dl for the low control test and 218 mg/dl for the high control test, which fall within the correct range for the low control and the high control.The complaint was not confirmed.The issue reported was likely caused by improper storage of test strips that led to contamination of the strips; the sample functioned as intended, and medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|