| Model Number |
MAJ-210 |
| Medical Device Problem Codes |
Fracture (1260); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported that during a diagnostic bronchoalveolar lavage and ultrasound examination, while the patient was under sedation, the rubber portion of the single use biopsy valve was chipped in a half-moon shape.The dropped-off portion was searched for, but its whereabouts are unknown.The customer also confirmed that an attempt was made to retrieve the broken portion, but no further details were provided.The procedure was completed with a similar device.There were no further reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection, and the reported failure was confirmed.The biopsy valve was found to be partially missing during the device evaluation.Based on the results of the investigation, a root cause could not be identified.It is possible that inserting and removing the syringe at an angle placed stress on the slit inside the biopsy valve, causing it to fracture.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|