• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA 4K UHD CAMERA CONTROL UNIT; VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA 4K UHD CAMERA CONTROL UNIT; VIDEO SYSTEM CENTER Back to Search Results
Model Number OTV-S400
Medical Device Problem Codes Communication or Transmission Problem (2896); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the video system center front panel liquid crystal display and the screen did not proceed to the main screen during inspection for use.During the initial evaluation of the device error code e116 was observed.There was no patient harm reported.
 
Additional Manufacturer Narrative
The investigation is still ongoing.A supplemental report will be submitted when the investigation has been completed or when new and significant information becomes available.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA 4K UHD CAMERA CONTROL UNIT
Common Device Name
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50923510
Report Number3002808148-2026-34054
Device Sequence Number5776719
Product Code FET
UDI-Device Identifier04953170374630
UDI-Public04953170374630
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K151011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Type of Report Initial
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOTV-S400
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/07/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-