| Catalog Number |
368607 |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information:\
d.9: device available for evaluation: yes\
h.3: device eval by manufacturer? yes\
d.9: returned to manufacturer on: 01-oct-2026\
h.1 imdrf annex a grid (1): a040102 - loss of or failure to bond (1068)\
\
investigation summary:\
bd received 10 samples and 4 photos for investigation.The ten returned samples were subjected to a visual functional test for defective locking mechanism and hub/collar separation.All samples passed testing, as there were no signs of damage to the device or separation during activation of the safety shield.Additionally, four customer photos were received, with one indicating batch details on the shelf carton and three images indicating hub/collar separation.\
device history record review confirmed that product quality was evaluated during manufacturing with prescribed inspections, all of which were in compliance with requirements.No related quality notifications or nonconformance reports were associated.\
this complaint has been confirmed for the indicated failure mode: hub/collar separation.\
no definitive root cause could be established for the reported defect.\
complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
Report 3 of 4\
\
it was reported that when using the bd vacutainer® eclipse¿ blood collection needle, an unspecified number of units exhibited hub/collar separation when attaching the holder or when attempting to activate the safety mechanism.No adverse impact was reported regarding the patient or user.
|
| |
|